Type:CX-IAC-1401A industrial acid / alkali concentration meter

Illustrate:It’s mainly used for monitoring the concentration of the acid / alkali regenerated by ion exchange resin in power plants and chemical industry.

CX-IAC-1401A Industrial Acid / alkali Concentration Meter


Brief Introduction
1.It is a kind of on-line instrument that is equipped with a microprocessor
2.Used for water quality analysis and monitoring. 

Main Features
1.Large vacuum fluorescent display (VFD) and man-machine interaction function
2.It is available with automatic temperature compensation and alarm discrimination.
3.Available with history curve, real-time trend chart and clock.
4.There are four kinds of recorders. It is available with isolated current output.
5.The user can set the password or consult the “Help” menu or the “Service Guide” Menu.
6.It has power-off protection. The data can be saved for 5 years under a continuous power-failure condition.
7.The communication interface can be connected directly to a computer and enables remote operation and control.
8.The instrument adopts universal module power supply.
9.It is equipped with acid / alkali sensor electrode. It can be easily used and maintained.

Main Parameters

Measuring range and error 0.0-10.0%; temperature: 0-60°C
Intrinsic error of the whole instrument Concentration: ±0.5%; temperature: ±0.5°C
Intrinsic error of the electronic unit concentration: ±0.25% ±l unit; temperature: ±0.5°C
Repeatability ±0.25% (concentration); ±1 Unit; ±0.25%(concentration);±1 unit
Stability ±0.25% (concentration); ±1 unit / 24 h
Automatic temperature compensation range 0~60 (with 25°C as the reference temperature)
Automatic temperature compensation error Concentration: ±0.25%
Output current error ≤±l.0%FS
Maximum output load and error For 0-10mA, the maximum output load is l.5KΩ: ±1%F.S.
For 4-20mA, the maximum output load is 750Ω±1%F.S
Normally-open contact of the alarm relay 3A 240VAC, 6A 28VDC or 120VAC
Network output Adopted balanced differential transmission without any repeater.
The maximum transmission distance is 1.2km.The maximum differential transmission with a repeater is 10.0km.
Water sample pressure 0.6Mpa
Alarm error of the electronic unit ±0.2% (concentration)
Data storage capacity 55 hours (1 point / 10s); 14 days (1 point / 1min); 68 days (1 point / 5min)
Saving time of data under a continuous power-failure condition 6 years
Power supply 80~265VAC 50 ±HZ
Overall dimension of the instrument Electronic unit: 151 (length) x 151 (width) x 260 (depth) mm
Dimension of the hole 138 x 138mm
Working conditions ambient temperature:0-60°C; relative humidity should be 90% or less.
Connection of the inlet and outlet water tubes Φ6, Φ8, Φl0 and Φl2

Last:CX-IAC-1401 industrial acid con.. Next:Without the