Type:CX-ICM-1101 industrial conductivity meter

Illustrate:Widely used for continuous monitoring of the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy, environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.

CX-ICM-1101 industrial conductivity meter

Brief Introduction
1.It is our company’s latest top-class instrument equipped with a microprocessor
2.It is characterized by its intelligent operation, multi-function and high measuring performance and its strong environmental adaptability
 
Main Features
1.Display:conductivity, output current, temperature, time and status.
2.Bitmap type liquid crystal display modulewith high resolution is adopted.
3.Thereis no symbol or code that is defined by the manufacturer.
 
Main Parameters

Measuring range 0.01~20μS/cm(Electrode:K=0.01),00.1~200μS/cm
(Electrode: K=0.1)
1.0~2000μS/cm(Electrode:K=1.0),10~20000μS/cm
(Electrode:K=10.0)                       
30~600.0mS/cm  (Electrode: K=30.0)
Intrinsic error of the electronic unit 2. : conductivity: ±0.5%FS, temperature: ±0.30C
Automatic temperature compensation 0~99.9C, with 25C as the reference temperature
Water sample tested 0~99.9C, 0.6MPa
Intrinsic error of the instrument Conductivity: ±1.0%FS, temperature: ±0.5C
Automatic temperature compensation error ±0.5%FS
Repeatability error ±0.2%FS±1 Unit
Stability ±0.2%FS±1 unit/24h
Isolated current output 0~10mA ( load<1.5kΩ),4~20mA (load<750Ω)
Output current error ≤±l%FS
Error caused by ambient temperature ≤±0.5%FS
Error caused by the supply voltage  ≤±0.3%FS
Alarm relay  AC 220V, 3A
Communication interface RS485 or 232 (optional)
Power supply AC 220V±22V, 50Hz±1Hz
Protection grade IP65
Clock accuracy ±1 minute/month
Data storage capacity l month (1 point/5 minutes)
The time for Saving data 10 years
Overall dimension 146 (length) x 146 (width) x 108 (depth) mm;
Dimension of the hole 138 x 138mm
Working conditions ambient temperature0~60C; relative humidity <85%
Note: The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, and 30.0.

Last:CX-ICM-1101A industrial conduct.. Next:CX-ICM-1104 industrial conducti..

Contact Us

Office Address: Room 402, No. 650, Xinzhuan Rd., Jiuting Town, Songjiang District, Shanghai, China (Mainland)

Phone:021-57635022

E-mail:info@gnflowmeter.com

E-mail:sales@gninstruments.com

關閉
主站蜘蛛池模板: 资源县| 中宁县| 庆城县| 团风县| 无极县| 天水市| 金山区| 巫山县| 弥渡县| 炉霍县| 塔城市| 通城县| 洪湖市| 邛崃市| 富裕县| 南雄市| 那曲县| 乌拉特后旗| 长沙县| 巴林右旗| 临西县| 宜黄县| 鹤岗市| 星子县| 五指山市| 武安市| 长丰县| 腾冲县| 台江县| 石棉县| 深州市| 东兰县| 盘山县| 阜南县| 奉化市| 武定县| 枣庄市| 张家界市| 孙吴县| 镇宁| 天镇县|